ATR correction
The ATR correction is only used with IR spectra, that have been recorded with a ATR accessory. In this case, data points must be corrected for a relative shift introduced by the instrument configuration.
The results of an ATR correction strongly depend on a good base line of the spectrum. So please perform a baseline correction first to obtain good results.
An ATR correction is performed as follows:
ATR correction menu command
To perform an ATR correction using the menu command, please follow the instructions below:
-
Activate the spectrum, you like to correct.
-
From the Mathematics menu, select the ATR Correction command.
-
Enter a reference wave number into the Reference Wave Number field.
-
Press the Calculate button.
ATR correction keyboard shortcut
None.
Advanced ATR correction menu command
To perform an advanced ATR correction using the menu command, please follow the instructions below:
-
Activate the spectrum you'd like to correct.
-
Choose Advanced ATR Correction from the Mathematics menu.
-
Enter the refractive indexes of the sample and the crystal.
-
Enter the incident angle of the measurement.
-
Either enter values for the absorption coeffizient and the electric field of the evanescent wave or calculate a machine constant used for further ATR corrections by performing a calculation on a spectrum with a reference. To recalculate the constant, give a new reference spectrum.
-
Press the Calculate button.
Advanced ATR correction keyboard shortcuts
None.